If you are redirected to this guide, it indicates that your 3rd mirror vertical path cannot be aligned, or you have replaced the laser tube, the 2nd mirror seat assembly, etc.
In this guide, the following advanced adjustments will be covered.
The horizontal line for the path from the 2nd mirror to the 3rd mirror entrance.
The vertical line for the path from the 2nd mirror to the 3rd mirror entrance.
The vertical path from the 3rd mirror to the bed.
Before you start
Please make sure your 1st to 2nd mirror path and the 2nd to 3rd mirror path are well aligned. Otherwise, the steps on this page will make your optical path worse.
Please refer to this article, 4-10 Optical Path Adjustment, from Step 1 to Step 7.1 to ensure the mentioned paths are well aligned.
Step 1
Please prepare the following tools.
OPA Tool
Double-head wrench
Pliers
Double-sided tape
2.5mm Allen key
Step 2
Use the OPA tool to verify that the dot is positioned at the center of the 3rd mirror's entrance. This step will teach you to verify the beam's vertical alignment at the 3rd mirror entrance using an OPA tool.
Please use the link to download the file and use a 3mm transparent acrylic to engrave and cut.
The tool will be put on the screw of the 3rd mirror entrance. The longer line at the left will be alinged with the edge of the metal piece.
If your machine can't cut through any area of the bed, you have to make a temporary tool. Please refer to the guide, Temporary OPA Tool Guide.
The tool will be put on the screw of the 3rd mirror entrance. The "x" mark side will be on the left side and the 10mm line will be alinged with the metal piece. Make sure the bottom of the tool will be sitting on the top surface of the screw.
Once you have the tool, move the laser head to the center and make 5 pulses. Use the tool to check if the dot is close to the vertical line.
Check the vertical line with the tool. Align the tool's circle or points with the entrance and check the dot's position.
Acrylic
Good Example
Bad Example
Tape
Good - Center
Bad - Not Center
Determine your result and the next action to take:
Recheck the optical path from the 2nd mirror to the third mirror to see if the dots overlap. Please refer to Step 7 in other article. If the dots don't overlap, please go through Step 7.1.
Go to Step 3 after the 2nd to 3rd mirror alignment.
Step 3
Check the horizontal line using the OPA tool.
Once you have the tool, move the laser head to the center and make 5 pulses. Use the tool to check if the dot is close to the horizontal line.
Check the vertical line with the tool. Align the tool's circle or points with the entrance and check the dot's position.
The dot should be centered on the line. It’s okay if it’s slightly above or below, as long as it touches the line. If it doesn’t touch the line, it’s not acceptable.
Acrylic
Good Example
Acceptable Example
Bad Example
Tape
Good Example
Acceptable Example
Bad Example
Determine your result and the next action to take:
Reset and align the path from the 3rd mirror to the entrance of the focus lens.
Firstly, loosen the nuts on the 3rd mirror and rotate them counterclockwise to reset the 3rd mirror alingment.
Place a two-layer tape on the righ of the nozzle. Make two pulses to create a dot.
Adjust the screws to make the dot be in the center of the ring.
Demonstration
(The video is an actual operation from the FLUX staff. It is a long video, you can change the speed, but the concept and the process is the key for you to understand how to adjust. Please enable the subtitle.)